Experimental workflow supports troubleshooting in label-free biosensor assay development, indicating practical tools may enhance success rates.
Label-free biosensors offer powerful platforms for detecting molecular interactions, but developing robust assays on these systems presents several challenges due to the complexity of the testing systems and intersection of disciplines. In this work, we describe a lightweight, 3-step experimental workflow supporting troubleshooting and root cause analysis that we developed during the design and optimization of silicon photonic biosensor assays in an academic research setting. Because such environments often lack the resources and formal quality-management infrastructure available in industry, our approach emphasizes practicality and ease of adoption. Drawing from our own assay failures and successes, we identify common failure modes, propose structured troubleshooting workflows, and provide case studies illustrating the application of established frameworks, including the Five Whys, the Plan-Do-Check-Act cycle, Open-Narrow-Close and the Ishikawa (fishbone) diagram. By applying this framework to our research, we increased our assay yield by a factor of 2, from ~45% to ~90%. This paper aims to support other teams engaged in label-free assay development by equipping them with practical tools and experimental best practices.
No takes yet. Share an insight, caveat, or question.
Chowdhury et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: