This evaluation quantifies operator variability in ultrasound assessments for DDH, revealing potential in AI-assisted measurements for greater consistency.
Ultrasound examination using the Graf method is widely applied for early detection of developmental dysplasia of the hip (DDH), but intra- and inter-operator variability remains a limitation. This study aimed to quantify operator variability in hip ultrasound assessments and to validate an AI-assisted system for automated α-angle measurement to improve reproducibility. Thirty participants of different experience levels, including trained clinicians, residents, and medical students, each performed six ultrasound scans on a standardized infant hip phantom. Examination time, iliac margin inclination, and α-angle measurements were analyzed to assess intra- and inter-operator variability. In parallel, an AI-based system was developed to automatically detect anatomical landmarks and calculate α-angles from static images and dynamic video sequences. Validation was conducted using the phantom model with a known α-angle of 70°. Clinicians achieved shorter examination times and higher reproducibility than residents and students, with manual measurements systematically underestimating the reference α-angle. Static AI produced closer estimates with greater variability, whereas dynamic AI achieved the highest accuracy (mean 69.2°) and consistency with narrower limits of agreement than manual measurements. These findings confirm substantial operator variability and demonstrate that AI-assisted dynamic ultrasound analysis can improve reproducibility and reliability in routine DDH screening.
No takes yet. Share an insight, caveat, or question.
Ohashi et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: