Newly developed optical frequency comb shows a repetition rate of 100 MHz, highlighting cost-efficient design improvements.
We designed and built an adjustable repetition rate optical frequency comb using off-the-shelf components with several improvements to an established erbium fiber design. This design, built fully by an undergraduate, was assembled at a fraction of the cost of a commercial comb. This comb has a repetition rate of roughly 100 MHz with a tuning range of 3 MHz. A stabilization control loop for fᵣₑₚ achieved long term drifts of < 1 Hz over several days. This comb will eventually be implemented for use in dual-comb spectroscopy.
No takes yet. Share an insight, caveat, or question.
Latchford et al. (2025) studied this question.