This immunoassay demonstrates high sensitivity and specificity for detecting galactomannan in patients, highlighting its diagnostic potential.
Introduction Invasive pulmonary aspergillosis (IPA) poses significant diagnostic challenges in immunocompromised patients, with galactomannan (GM) detection serving as a key biomarker. Existing Platelia™ GM enzyme-linked immunosorbent assay (ELISA) faces limitations in throughput and turnaround time. Methods We developed a magnetic bead-based chemiluminescent enzyme immunoassay (CLEIA) for GM detection in serum/bronchoalveolar lavage fluid (BALF). Results Optimized CLEIA parameters enabled 30-min analysis versus 120-min for ELISA. Validation with 241 clinical specimens demonstrated robust analytical performance: diagnostic thresholds of ≥ 0.20 CLEIA Units for serum and BALF matrices (ROC-AUC 0.87, p<0.0001), analytical sensitivity of 0.50 ng/mL, and precision with ≤14.40% total CV, and no hook effect ≤200 ng/mL. Method comparison against the Platelia ELISA revealed 86.72% overall diagnostic agreement (Cohen’s k=0.72), while cross-reactivity analyses indicated specificity limitations exclusively with Histoplasma -positive specimens, consistent with parallel GM ELISA results. Discussion This assay enables 75% processing time reduction with uncompromised diagnostic accuracy, positioning it as a high-throughput alternative for rapid IA screening in high-risk cohorts.