Retrospective review indicates racial disparities in cerebral near infrared spectroscopy accuracy in infants, highlighting potential measurement bias.
Objective: Pulse oximeters overestimate arterial oxygen saturations in black versus white adults, children, and infants. While race's impact on near-infrared spectroscopy (NIRS) accuracy is less studied, some adult research suggests decreased accuracy in black patients. This study investigates the effect of race on NIRS accuracy in infants in a cardiac intensive care unit (CICU). Study Design: A retrospective chart review was conducted for infants admitted to St. Louis Children’s Hospital CICU from 2017-2023. Bland-Altman plots, Pearson correlations, and mean biases were analyzed. Result: 254 infants (13% Black, 87% White) provided 3,687 central venous oxygen saturation (ScvO2)-cerebral regional oxygen saturation (rScO2) pairs. Measurement bias was -3.2% in Black infants and +0.1% in White infants (p<0.01). Conclusion: Cerebral NIRS underestimates ScvO₂ in Black infants but maintains negligible measurement bias in White infants. This is the first study to assess race and NIRS accuracy in infants; the difference is statistically significant but not clinically relevant in most contexts.
No takes yet. Share an insight, caveat, or question.
Marshall et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: