Abstract We demonstrate the use of machine-learning aided UV absorbance spectroscopy to detect contamination during cell therapy product (CTP) manufacturing. The proposed method leverages a one-class support vector machine model to analyse the absorbance spectra of cell cultures and predict if a sample is sterile or contaminated. This label-free and non-invasive method provides a rapid output (30 minutes) with minimal sample preparation and sample volume (<1mL). Spiking of 7 bacterial species into mesenchymal stromal cells supernatant aliquots from 6 commercial donors showed that contamination events could be detected at low inoculums with mean true positive and negative rates of 92.7% and 77.7% respectively. Accounting for donor variation improved specificity to 92.0%. In cells spiked with 10 CFUs/mL of E.coli, contamination was detected at the 21-hour timepoint, demonstrating comparable sensitivity to compendial USP <71> test (~24 hours). We hypothesize that spectral differences between nicotinic acid and nicotinamide in the UV region are the underlying mechanisms for contamination detection. This technique can be deployed as a preliminary test during different CTP manufacturing stages where cultures can be monitored real-time to detect early onset of microbial contamination, assuring safety of CTP prior to infusion into patients.
No takes yet. Share an insight, caveat, or question.
Chelvam et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: