Key result
Cardiac CT yields substantial reader agreement for post-Watchman FLX thickening and leak with kappa over 0.61.
Why the study?
Is the interpretation of hypoattenuated thickening and peridevice leak on cardiac CT following Watchman FLX LAAC reproducible between experienced and novice raters?
Observational (n=100)
Blinded to prior evaluations
Yes
Is the interpretation of hypoattenuated thickening and peridevice leak on cardiac CT following Watchman FLX LAAC reproducible between experienced and novice raters?
Effect estimate: kappa ≥0.61
Cardiac CT interpretation of hypoattenuated thickening and peridevice leak following Watchman FLX LAAC shows high inter- and intra-rater reproducibility even with minimal training.
Aims To assess the reproducibility of interpreting hypoattenuated thickening (HAT) and peridevice leak (PDL) using cardiac computed tomography (CT) imaging following Watchman FLX left atrial appendage closure (LAAC) Methods In this multicentre retrospective reproducibility study, 100 anonymized post-LAAC cardiac CT scans were evaluated within and results the same cardiac phase by an experienced and a novice rater blinded to prior evaluations. All scans were evaluated twice by each rater, assessing overall HAT and PDL categories as well as specific associated findings based on suggested algorithms for post-LAAC interpretation. Inter- and intra-rater agreement and reliability were evaluated using absolute agreement, Cohen’s kappa, and Kendall’s tau for categorical variables, and mean difference, Bland–Altman plots, limits of agreement (lower and upper), and intra-class correlation coefficients (ICCs) for continuous variables. Within overall categories of both HAT and PDL, substantial agreement (kappa ≥0.61) and reliability (Kendall’s tau-b ≥ 0.75) were observed. Specifically, identifying high-grade HAT (kappa ≥0.78) and distal patency (kappa ≥0.85) displayed the highest agreement within HAT and PDL interpretation. Meanwhile, measuring the height of the proximal screw hub cove represented the least reliable HAT assessment among both inter- and intra-rater comparisons (ICC < 0.75), while suspected leak mechanism represented the least reproducible PDL measure Conclusion Despite only minimal training of one rater, overall high levels of inter- and intra-rater agreement and reliability were observed across the chosen algorithms for interpretation of HAT and PDL following Watchman FLX LAAC. The prognostic implications of the included variables are to be explored in future trials and registries.
No takes yet. Share an insight, caveat, or question.
Kramer et al. (2024) conducted an observational in Post-Watchman FLX left atrial appendage closure (n=100). Cardiac CT interpretation of hypoattenuated thickening and peridevice leak was evaluated on Inter- and intra-rater agreement and reliability for overall HAT and PDL categories (kappa ≥0.61). Cardiac CT interpretation of hypoattenuated thickening and peridevice leak following Watchman FLX LAAC demonstrated substantial inter- and intra-rater agreement (kappa ≥0.61).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: