Several false positive low serum cryptococcal antigen (SCrAg) reports by lateral flow assay (LFA) were identified in late 2016 at our tertiary care hospital. After the recall and correction of the problem in the reagent, we studied the significance of SCrAg LFA ≤ 1:10 from January 2017 to October 2023. Of 20 patients with 31 samples of SCrAg LFA ≤ 1:10, 14 patients (70%) were classified as true positives, four (20%) were indeterminate, and only two (10%) patients were false positives. If a new SCrAg LFA ≤ 1:10 is detected, it should be repeated, and additional workup should be pursued.