The DR.BEAT project aims to develop an accelerometer-based, wearable sensor system for measuring ballistocardiographic (BCG) signals, coupled with signal processing and visualization, to support cardiac health monitoring. A rule-based heartbeat detection was developed to enable the derivation of health parameters independent of an existing reference. This paper outlines the algorithm’s methodology and provides an initial evaluation of its performance based on seismocardiographic (SCG) measurements obtained from an initial study involving twelve heart-healthy adults. On average, 87.6% of the heartbeats over all measurements, 97.6% of the heartbeats at rest and 71.9% of the heartbeats during physical stress could be detected.
No takes yet. Share an insight, caveat, or question.
Wolf et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: