It has been challenging to achieve ultrashort echo times with two-dimensional acquisitions. However, the use of a prepulse and a whole volume hard excitation may address these limitations. In addition, such a pulse sequence would allow for the use of simultaneous multi-slice pulses to reduce scan times. We aim to design a low specific absorption rate, short duration MultiPINS prepulse that excites a slice profile with sharp, narrow slice gaps. Optimization suggests that a pulse with time bandwidth product (TBW) of 49.43 is optimal, but further investigation suggests that a pulse with TBW as low as 36 may be feasible.