Noise in computed tomography (CT) is inevitably generated, which lowers the accuracy of disease diagnosis. The non-local means approach, a software technique for reducing noise, is widely used in medical imaging. In this study, we propose a noise reduction algorithm based on fast non-local means (FNLMs) and apply it to CT images of a phantom created using 3D printing technology. The self-produced phantom was manufactured using filaments with similar density to human brain tissues. To quantitatively evaluate image quality, the contrast-to-noise ratio (CNR), coefficient of variation (COV), and normalized noise power spectrum (NNPS) were calculated. The results demonstrate that the optimized smoothing factors of FNLMs are 0.08, 0.16, 0.22, 0.25, and 0.32 at 0.001, 0.005, 0.01, 0.05, and 0.1 of noise intensities, respectively. In addition, we compared the optimized FNLMs with noisy, local filters and total variation algorithms. As a result, FNLMs showed superior performance compared to various denoising techniques. Particularly, comparing the optimized FNLMs to the noisy images, the CNR improved by 6.53 to 16.34 times, COV improved by 6.55 to 18.28 times, and the NNPS improved by 10