Modern PET scanners offer precise TOF information, improving the SNR of the reconstructed images. Timing calibrations are performed to reduce the worsening effects of the system components and provide valuable TOF information. Traditional calibration procedures often provide static or linear corrections, with the drawback that higher-order skews or event-to-event corrections are not addressed. Novel research demonstrated significant improvements in the reachable timing resolutions when combining conventional calibration approaches with machine learning, with the disadvantage of extensive calibration times infeasible for a clinical application. In this work, we made the first steps towards an in-system application and analyzed the effects of varying data sparsity on a machine learning timing calibration, aiming to accelerate the calibration time. Furthermore, we demonstrated the versatility of our calibration concept by applying the procedure for the first time to analog readout technology.
No takes yet. Share an insight, caveat, or question.
Naunheim et al. (2024) studied this question.