This paper describes in detail the RP-HPLC method used to evaluate the content of evinacumab (EVC) in combination with evkeeza injections. The mobile phase used in the assay is a mixture of potassium dihydrogen phosphate and methanol in an isocratic ratio of 70:30 (v/v) on a column Shimpac solar C18, 150*4.6, 5μm column. The flow rate was 1.0 ml/min, isocratic. At 226 nm, the quantification and detection of EVC were done concurrently. It was discovered that the response obtained was linear over the concentration range of 75-225 μg/ml, and the estimated RT in minutes for EVC was 4.063. 1.805 μg/ml was the LOD and 6.017 μg/ml was the LOQ. For the EVC assay, precision was 0.3% RSD and accuracy was 100%.