In this dual-center study, we assessed the BioHermes A1C EXP M13 system for point-of-care (POC) HbA1c testing against two NGSP-certified HPLC instruments, the Bio-Rad D100 and Tosoh G8. Analyzing 605 samples, we evaluated the A1C EXP's reproducibility, sensitivity, specificity and impact of anemia on HbA1c measurements. The device showed excellent reproducibility with CVs under 2.4% and high sensitivity and specificity for diabetes diagnosis-98.1% and 96.8% against D100, and 97.1% and 96.7% against G8. Passing-Bablok regression confirmed a close correlation between A1C EXP and the HPLC instruments, with equations