In the near-infrared wavelength regime, atmospheric turbulence fluctuates at a scale of a few milliseconds, and its precise control requires the use of extreme adaptive optics (XAO) systems equipped with fast and sensitive detectors operating at kHz speeds. The C-RED One cameras developed by First Light Imaging (FLI), based on SAPHIRA detectors made of HgCdTe e-APD array sensitive to 0.8-2.5 μm light, featuring a 320x256 pixels with 24 μm pitch, offering sub-electron readout noise and the ability to read subarrays, at frame-rates of up to few 10-kHz, are state-of-the-art for XAO wavefront sensing. The Observatory of Geneva purchased two C-RED One cameras identified as necessary for RISTRETTO (a proposed high-contrast high-resolution spectrograph for the VLT) and SAXO+ (an upgrade of the VLT/SPHERE XAO system) projects. We present a comprehensive characterization and comparative analysis of both the cameras. We present test results examining key noise contributors, including readout noise, detector bias, etc. And we also study their temporal variability. Additionally, we assess the conversion gain and the avalanche gain calibration of the detector. We also study the evolution some of these parameters over time.
No takes yet. Share an insight, caveat, or question.
Shinde et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: