Photomultiplier tubes (PMTs) are widely used in particle experiments for photon detection. PMT waveform analysis is crucial for high-precision measurement of the position and energy of incident particles in liquid scintillator (LS) detectors. A key factor contributing to the energy resolution in large liquid scintillator detectors with PMTs is the charge smearing of PMTs. This paper presents a machine-learning-based photon counting method for PMT waveforms and its application to the energy reconstruction, using the JUNO experiment as an example. The results indicate that leveraging the photon counting information from the machine learning model can partially mitigate the impact of PMT charge smearing and lead to a relative 2.0% to 2.8% improvement on the energy resolution at different energies.
No takes yet. Share an insight, caveat, or question.
Jiang et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: