The handheld optical imager utilizes 660 nm and 940 nm dual wavelength LEDs and photodetectors. There are four sources, each followed by three detectors and arranged equidistant from one another. A switching circuit was implemented using a 555 timer in astable mode to facilitate the switching of the LED at the aforementioned wavelengths. To acquire data, the rectangular-shaped scanner was kept on a volunteer's palm and moved towards the tip of the finger. At each position of the imager, red and near-IR radiations were sent through the tissue, and the detectors received backscattered radiation. The current signals obtained from the photodiodes were converted into voltages using a trans-impedance amplifier and read into the computer using NI-DAQ in twelve parallel channels. Using Beer-Lamberts law, the attenuation coefficient was calculated and displayed as images. These images show the difference between bony structures and other tissue in the fingers. The distal and proximal phalangeal joints appear lower in intensity in comparison to the bony structure of the finger. The oxygen saturation map of the scanned area was also obtained. Subjects with arthritic conditions have inflamed joints. The arthritic joints are sensitive to changes in oxygen saturation. Therefore, the imager could distinguish between normal and inflamed joints and be a screening tool for arthritis. The FWHM values are calculated to validate the obtained results.