This study explores the advantages of a dual axial flux permanent magnet synchronous machine (AFPMSM) structure compared with a single AFPMSM, addressing the increasing demand for devices with high torque and compact dimensions for a left ventricular assist device (LVAD) pump. The AFPMSM design offers superior torque and power densities, making it a potential solution to the challenges posed by the need for compact yet powerful electric machines. The investigation involves the comparison of efficiency, torque ripple, and thermal losses between a dual-motor AFPMSM configuration and a single-motor setup. Key findings include a significant reduction in torque ripple, from 22% to 12.6%, achieved by assembling dual AFPMSMs at different angles. The dual-motor system also demonstrates the ability to maintain a high efficiency of approximately 60% through optimal load sharing, even at high loads. Thermal analysis revealed substantial temperature reductions, making the dual AFPMSM structure particularly advantageous, especially for small-scale applications. Overall, this research highlights the multifaceted benefits of employing a dual AFPMSM structure in addressing the complexities of high torque and small size requirements in electric machine design.
No takes yet. Share an insight, caveat, or question.
Karabulut et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: