Our research highlights the potential of Diffuse Reflectance Spectroscopy (DRS) in detecting cortical breaches during pedicle screw placement. We propose a sideways-looking fiber-optic probe, integrating diffuse light emission with both forward and sideways light collection. Experiments on an optical tissue phantom validate the probe's potential to distinguish bone tissues and provide real-time guidance for spine surgery. Our findings prove that DRS with diffuse emission can detect perpendicular breaches, and demonstrate how the integration of a 45° slanted fiber coated with gold enables parallel breach detection, advancing spine surgery by allowing for accurate pedicle screw placement.
No takes yet. Share an insight, caveat, or question.
Losch et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: