Employing an automated monitoring system (AMS) for data acquisition offers benefits, such as reducing the workload, in the kinetic study of suspended photocatalytic batch reactions. However, the current methods in this field tend to narrowly focus on the substrate and often overlook the optical characteristics of both the mixture and solid particles. To address this limitation, in this study, we propose a novel AMS based on online circulatory spectrophotometry (OCS) and incorporate debubbling, aeration, and segmented flow (DAS), named DAS-OCS-AMS. Initially, a debubbler is introduced to mitigate the issue of signal noise caused by bubbles (SNB). Subsequently, an aerated and segmented device is developed to address the issue of particle deposition on the inner wall of the pipeline (PDP) and on the windows of the flow cell (PDW). The proposed DAS-OCS-AMS is applied to monitor the kinetics of the photocatalytic degradation of Acid Orange Ⅱ by TiO