A versatile vapor generation module has been developed for the purpose of both field water vapor isotope calibrations and laboratory liquid water isotope measurements. The vapor generation module is fully scalable allowing in principle an unlimited number of standards or samples to be connected, opening up the possibility for calibrating with multiple standards during field deployment. Compared to a standard autosampler system, the vapor generation module has a more than 2 times lower memory effect. The vapor generation module can in principle generate a constant stream of vapor with constant isotopic composition indefinitely. We document an Allan Deviation for 17O-excess (Δ17O) of less than 2 per meg for an approximate 3 hour averaging time. For similar averaging time the Allan Deviation for