Introduction: Permanent pacemaker implantation (PPI) in neonates is challenging with respect to indications, device selection, implantation technique, and long-term outcomes. Complex anatomy, the need for long-term pacing with high rates, and a problematic postoperative period are the major problems. Methods: We prospectively followed up 22 newborns who underwent PPI below 28 days of life at our institute. Results: The median age at implantation was 2 days (interquartile range 1–9 days), and 9% were born preterm. The average heart rate before implantation was 46.4 ± 7.2 bpm. Maternal lupus antibodies were positive in 8 (36.4%) neonates, whereas 11 (50.0%) had associated congenital heart disease. Nineteen neonates underwent single chamber (VVI) and three underwent dual chamber (DDD) pacemaker implantation. Over a median follow-up of 46 months (range 2–123 months), the average ventricular pacing percentage was 87.5 ± 24.9%, with a stable pacing threshold. Seven children underwent pulse generator replacement due to battery depletion at a median age of 47 months. Pacing-induced ventricular dysfunction was seen in five children at a median age of 23.6 months, and two underwent upgradation to cardiac resynchronization therapy. Overall mortality was 13.6%, all due to tissue hypoperfusion and lactic acidosis in the postimplantation period. Conclusions: PPI in neonates has a favorable outcome with excellent lead survival. Overall mortality is 13.6%, which is predominantly in the postimplantation period and related to myocardial dysfunction.