A portable electrochemical system based on rGO/AuNPs/Nafion was constructed for selenium (IV) field determination in coastal seawater of the North Yellow Sea, East China Sea and South China Sea. This system demonstrated a low determination limit (0.03 μg/L), a wide linear range of concentration (0.1–30 μg/L), exceptional reproducibility and repeatability, and excellent selectivity. Each individual, entire test could be accomplished within 8 min. In Yantai Sishili Bay, the Se (IV) concentrations ranged from 0.16–3.59 μg/L (average 1.43 μg/L). In Shengshan Island, the Se (IV) concentrations ranged from 0.20–5.18 μg/L (average 2.06 μg/L). In Zhelin Bay, the Se (IV) concentrations ranged from 0.87–4.87 μg/L (average 3.13 μg/L). The average concentrations of Se (IV) in surface seawater showcased an increasing trend from the north to the south of China, suggesting Zhelin Bay > Shengshan Island > Sishili Bay. The system develops a new, competitive method for water quality assessment, and contributes to the accurate study of the biogeochemical cycles of selenium.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: