Answers are grounded in this paper and cite their sources.
Stent-assisted coil embolization is effective for treating intracranial aneurysms, improving outcomes and reducing recurrence rates. However, accurately measuring the diameter of a previously placed stent during imaging can be challenging due to coil artifacts. This poses difficulties in determining the coil packing and size of additional stents needed during retreatment. In a reported case, the use of a balloon enabled precise assessment of stent deployment. A 50-year-old male with a history of basilar artery-left superior cerebellar artery aneurysm underwent coil embolization, direct clipping, and stent-assisted coil embolization (SAC) over a span of 14 years. However, the aneurysm showed reenlargement over time. To address the recurrence, a balloon was used to assess the previously placed Neuroform Atlas stent. Additional coils were inserted outside the stent, and a Low-profile Visualized Intraluminal Support Blue stent was added. Postoperatively, there were no new neurological issues, and a follow-up magnetic resonance imaging showed no ischemic lesions. Balloon-assisted stent visualization (BASV) may be a useful method in the retreatment of SAC. It has the potential to provide valuable information for treatment planning.
No takes yet. Share an insight, caveat, or question.
Sakamoto et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: