At the Heidelberg Ion-Beam Therapy Center (HIT), the track structure of carbon ions of therapeutic energy after penetrating layers of simulated tissue was investigated for the first time. Measurements were conducted with a primary beam energy of 3.5 GeV and polymethyl methacrylate (PMMA) absorbers of different thickness to realize different depths in the phantom along the pristine Bragg peak. Ionization cluster size (ICS) distributions resulting from the mixed radiation field behind the PMMA absorbers were measured using the PTB ion counting nanodosimeter. The data analysis revealed unexpectedly high mean ICS values compared to the stopping power calculations using the SRIM code and the data measured at lower energies within the BioQuaRT project. Subsequently, Geant4 simulations of the experimental setup were performed to characterize the composition of the mixed radiation field. The results indicate the potential presence of a marked contribution of nuclear fragments to the measured ICS distributions, especially at nonzero impact parameters.