Potentiometric sensor achieved a detection limit of 1.25 × 10−10 M for magnesium ions, indicating high selectivity and improved response times.
Magnesium (Mg2+) ions require sensitive and selective detection due to their low concentrations and coexistence with similar ions in matrices. This study developed a potentiometric ISE using a new modified polyurethane membrane. The membrane’s negative surface charge facilitates selective interaction with Mg2+ ion. Optimal performance was obtained at 0.0061% (w/w) κ-carrageenan and 0.0006% (w/w) D2EHPA. The ISE exhibited a near-Nernstian response of 29.49 ± 0.01 mV/decade across a 10−9–10−4 M concentration range (R2 = 0.992), with a detection limit of 1.25 × 10−10 M and a response time of 200 s. It remained stable in the pH range 6–8 for one month and demonstrated high selectivity over K+, Na+, and Ca2+ (Kij < 1). The repeatability and reproducibility tests yielded standard deviations of 0.15 and 0.39, while recovery rates confirmed analytical reliability. The water contact angle analysis showed a reduction from ~80° to ~69° after membrane conditioning, indicating increased hydrophilicity and improved interfacial for ion diffusion. FTIR analysis confirmed successful modification by reduced O–H peak intensity, while XRD verified the amorphous structure. SEM revealed a dense top layer with concave morphology, favorable for minimizing leakage and ensuring efficient ion transport within the sensing system.
No takes yet. Share an insight, caveat, or question.
Hanum et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: