Key result
DBS impedance revisions affect ~4% of patients, highest with older Medtronic systems and tremors.
Why the study?
Impedance abnormalities often cause a sudden loss of therapeutic effect in patients with deep brain stimulation, prompting this analysis across three major manufacturers to determine their incidence and causes.
What is the incidence and what are the primary causes of impedance abnormalities requiring revision surgery in patients with de novo deep brain stimulation?
Cohort (n=601)
No
What is the incidence and what are the primary causes of impedance abnormalities requiring revision surgery in patients with de novo deep brain stimulation?
Answers are grounded in this paper and cite their sources.
Impedance abnormalities requiring revision surgery in DBS patients are relatively rare (4.16%) and are most commonly caused by electrode-related issues, with variations observed across device manufacturers.
Background/Objectives: Patients with deep brain stimulation (DBS) require regular follow-up. When a sudden loss of therapeutic effect occurs, impedance abnormalities are often the underlying cause. If reprogramming cannot restore clinical benefit, revision surgery may be necessary to replace defective hardware. Since all three major manufacturers are used at our center, we analyzed our patient cohort to determine the incidence and causes of impedance abnormalities. Methods: All 601 patients who underwent de novo DBS implantation in Hamm between 2009 and 2025 were evaluated for impedance abnormalities. In cases requiring revision surgery, the specific cause was identified. The manufacturer, electrodes, and contacts involved were systematically analyzed. Results: A total of 25 of 601 patients required revision surgery. Revision rates were 2.67% in patients with Parkinson’s disease, 6.19% in those with a tremor, and 5.71% in those with dystonia. Across manufacturers, 7.6% of patients with a Medtronic system required revision surgery, compared with 3.4% of patients with an Abbott system and no patients with a Boston Scientific system. The primary causes of revision were electrode-related problems (19/25), followed by extension defects (6/25), connector issues (4/25), and, in one case, a generator defect (1/25). Conclusions: Only 4.16% of patients required revision surgery due to impedance abnormalities. Patients with a tremor and non-segmented electrodes showed a higher incidence than those with Parkinson’s disease or dystonia. Predominantly older Medtronic systems had the highest revision rate, whereas no Boston Scientific systems required revision. In most cases, the electrodes were the primary source of impedance abnormalities. A total of 52% of revisions were performed within two years and 92% were performed within six years of implantation.
No takes yet. Share an insight, caveat, or question.
Fortmann et al. (2026) conducted a cohort in Deep brain stimulation (Parkinson's disease, tremor, dystonia) (n=601). Deep brain stimulation systems was evaluated on Revision surgery due to impedance abnormalities. Among 601 patients with de novo deep brain stimulation, 4.16% required revision surgery due to impedance abnormalities, most commonly related to electrode problems.