February 5, 2026Radiation Protection Dosimetry2 citations

State-of-the-art 32 cm field-of-view digital PET/CT system: preliminary study for protocols optimization and DRLs update

View Full Paper
CFC. FulcheriOSOttavia SguazziniVPVeronica D’Ignoti Parenti

Key Points

  • The study aims to evaluate the performance of a new PET/CT system and update diagnostic reference levels for administered activity.
  • Evaluated the GE Omni Legend PET/CT using NEMA NU 2-2018 measurements and EARL2 accreditation tests.
  • Compared the results with the previous GE Discovery IQ scanner.
  • Conducted a retrospective analysis of standard-sized adult examinations to derive updated DRLs values.
  • Omni Legend showed improved sensitivity and noise equivalent count rate.
  • Achieved up to 60% reduction in injected activity while maintaining or improving image quality.
  • Diagnostic reference levels were lower than national and international benchmarks, indicating effective dose optimization.

Abstract

Abstract The aim of this study was to assess the impact of a state-of-the-art 32 cm axial field-of-view GE Omni Legend PET/CT system on administered activity and image quality, and to update Diagnostic Reference Levels (DRLs). System performance was evaluated using NEMA NU 2-2018 measurements and EARL2 accreditation tests. A comparison with the previous 5-ring GE Discovery IQ scanner was performed to analyze the administered activity, scan duration and image quality with and without Artificial Intelligence driven reconstruction. A retrospective analysis of standard-sized adult examinations was conducted to derive updated DRLs values. Omni Legend demonstrated improved sensitivity and Noise Equivalent Count Rate, enabling up to 60% reduction in injected activity while maintaining or improving image quality. Deep learning reconstruction enhanced image quality, particularly at lower administered activities. The DRLs were lower than the national and international reference levels, supporting dose optimization in clinical practice.