Key result
Normal V5-V6 T waves and terminal S waves on standard ECG help ascertain CSP capture.
Why the study?
Current criteria to ascertain proper lead placement in conduction system pacing require an EP system that is not widely available, while multiple ECG-only criteria remain unvalidated.
Do standard ECG criteria help ascertain conduction system pacing capture during implantation?
Cohort (n=482)
Do standard ECG criteria help ascertain conduction system pacing capture during implantation?
Standard ECG parameters, specifically V5-V6 normal T wave repolarization and terminal S waves, can serve as helpful tools to ascertain conduction system pacing capture during implantation without requiring an EP system.
Answers are grounded in this paper and cite their sources.
May aid conduction system pacing capture confirmation without EP systems; leaves open prospective validation.
Conduction System Pacing (CSP) is gaining acceptance when high ventricular pacing burden is expected, and as an alternative to cardiac resynchronization. Current accepted criteria to ascertain CSP proper lead placement requires an EP system, not widely available. There are multiple ECG only criteria not currently validated for CSP implantation. We extracted all CSP attempted implants. Various accepted criteria for CSP capture were evaluated for each patient, along with other ECG only criteria: T wave polarity in lateral leads (V5-V6); T wave polarity in D1, terminal S wave in lateral leads; terminal S wave in D1; and an initial r wave in V1. A cohort of 482 consecutive patients was evaluated, where 36% were women. Mean age was 77.3 y.o. (± 8.9 years; range 22-98 y.o.). Implant indications included: sick sinus (26.3%), AV block (34.9%), abnormal conduction induced cardiomyopathy (26.1%), AV node ablation (9.1%) and LBBB only (0.8%). Medtronic 3830 leads were mostly implanted. A His pacing was targeted in 5.2% of patients, and 51.4% had pre-implant narrow QRS. Leads were directed anteriorly, posteriorly and mid-septal in 28.3%, 18.6%, and 53.1% of cases, respectively (based on inferior lead polarities). A QRS (from the pacing spike) of 140ms or less was achieved in 60.0% of implants, a LVAT lesser than 75ms in 47.7% (lesser than 85ms in 78.1% of patients and impossible to evaluate in 6.0% of patients (due to Q-S in V6)). A V6-V1 interpeak interval greater than 33ms was achieved in 54.0% of patients (the criterion was impossible to measure in 36.3% of patients). When comparing possible ECG parameters which may guide CSP implants, V5-V6 normal T wave repolarization and terminal S waves in V5-V6 where found to be most helpful and provides another tool to ascertain CSP capture when r’/R’ are not seen (mostly in anterior oriented positions) or V6-V1 inter-peak is impossible to measure (lack of either positive V5-V6 or visible r’/R’ in V1).
No takes yet. Share an insight, caveat, or question.
Fakim et al. (2025) conducted a cohort in Conduction System Pacing (CSP) implants (n=482). Standard ECG criteria (V5-V6 T wave repolarization and terminal S waves) was evaluated on Evaluation of ECG parameters to guide CSP implants. Standard ECG parameters, specifically V5-V6 normal T wave repolarization and terminal S waves, were found to be helpful in ascertaining conduction system pacing capture.