Key result
Amplatzer PFO Occluder shows no difference in device syndrome versus Gore Cardioform in nickel hypersensitivity.
Why the study?
Does the Amplatzer PFO Occluder compared to the Gore Cardioform Septal Occluder reduce device syndrome in patients with nickel hypersensitivity undergoing PFO closure?
RCT (n=96)
Double-blind
randomized
Yes
Does the Amplatzer PFO Occluder compared to the Gore Cardioform Septal Occluder reduce device syndrome in patients with nickel hypersensitivity undergoing PFO closure?
Absolute Event Rate: 68.8% vs 75%
p-value: p=1.000
In patients with nickel hypersensitivity undergoing PFO closure, there was no significant difference in the incidence of device syndrome between the Amplatzer PFO Occluder and the Gore Cardioform Septal Occluder.
Background Although the lifetime stroke risk attributable to PFO is low, transcatheter PFO closure has demonstrated efficacy in secondary stroke prevention. Randomized trials have shown significant reductions in recurrent stroke risk compared to antithrombotic treatment alone. There are only two Food and Drug Administration and CE-approved devices for PFO closure: the Amplatzer PFO Occluder and the GoreCardioform Septal Occluder (GSO). Both devices are constructed from nitinol, an alloy of nickel and titanium. Nickel is a potent allergen responsible for type IV hypersensitivity in >20% of the population. While skin hypersensitivity, primarily expressed as contact dermatitis, is well-characterized, the impact of systemic exposure to nickel from implanted nitinol devices remains poorly understood. Purpose The aim of our analysis is to assess the potential influence of device selection in patients with known nickel hypersensitivity. Methods This is a post-hoc analysis of INSPIRE trial. Briefly, the INSPIRE (Investigation of Nickel Sensitization After Percutaneous Implantation of Patent Foramen Ovale Occluder) trial was an investigator-initiated, multicenter, prospective, double-blinded, randomized trial with blinded endpoint assessment, exploring whether patients with nickel hypersensitivity have an increased risk of adverse events following PFO closure. Nickel hypersensitivity was assessed using skin patch testing. The primary endpoint was device syndrome, a composite of patient-reported symptoms (chest pain, palpitations, migraines, dyspnea, and rash). In addition to the composite outcomes listed as the primary endpoint, the secondary endpoints included documented atrial arrhythmias [atrial fibrillation (AF), atrial flutter, or supraventricular tachycardia], major or minor bleeding, ischemic or hemorrhagic stroke, and all-cause mortality. Results Between January 2021 and September 2024, a total of 96 patients were included in the analysis and randomized to receive either the Amplatzer PFO Occluder (N = 48) or the GSO (N = 48). A total of 28 patients (29.2%) had been diagnosed with nickel hypersensitivity: 4 (4.2%) had a weak positive reaction, 16 (16.7%) had a strong positive reaction, and 8 (8.3%) had an extremely positive reaction. Sixteen patients were randomized to the Amplatzer PFO Occluder, and 12 patients were randomized to the GSO. No significant difference was observed in baseline characteristics. The incidence of the primary endpoint was comparable between the Amplatzer group (N = 11, 68.8%) and the GSO group (N = 9, 75.0%) (p = 1.000). No significant differences were observed in any secondary endpoints. Conclusions Our prospective, double-blinded, randomized trial with blinded endpoint assessment demonstrated that no significant differences were observed between the Amplatzer PFO Occluder and the GSO. Further larger studies are needed to provide more definitive insights into this issue.
No takes yet. Share an insight, caveat, or question.
Davlouros et al. (2025) conducted an RCT in Patent foramen ovale with nickel hypersensitivity (n=96). Amplatzer PFO Occluder vs. Gore Cardioform Septal Occluder (GSO) was evaluated on Device syndrome (composite of patient-reported symptoms: chest pain, palpitations, migraines, dyspnea, and rash) (p=1.000). The Amplatzer PFO Occluder resulted in a similar incidence of device syndrome compared to the Gore Cardioform Septal Occluder in patients with nickel hypersensitivity (68.8% vs 75.0%; p=1.000).