Multi-reader analysis compares image quality of different CT reconstruction methods for cochlear implants, guiding optimization.
PCCT reconstruction parameters substantially influence postoperative CI image quality. Ultra-high-resolution reconstructions provided the best combination of artifact suppression and fine structural detail, while iterative MAR and high-keV monoenergetic reconstructions performed the worst. These findings can guide reconstruction selection to optimize PCCT protocols for CI evaluation.
No takes yet. Share an insight, caveat, or question.
Dogra et al. (2026) studied this question.