February 11, 2026Bone & Joint Open0 citationsOpen Access

Comparing the express and enhanced workflows for restoration of hip length and combined offset using the Mako robotic arm-assisted total hip arthroplasty

TCTim CheokUniversity Medical Center GroningenVPVeronica PajnicQueensland HealthJVJulie F VermeirThe University of Queensland

Key Points

  • This study aims to compare the express and enhanced workflows of the Mako robotic system for total hip arthroplasty in restoring hip length and offset.
  • Retrospective cohort design comparing two workflows: enhanced vs. express.
  • Measured values were compared to the Mako derived values.
  • Bland-Altman plots and Theil's median slope were used for assessment.
  • Secondary outcomes included surgical time and postoperative complications.
  • Enhanced workflow showed a better median measurement for combined offset (1.50 mm vs. 3.13 mm).
  • No significant differences in hip length discrepancy or surgical time between groups.
  • Both workflows had no cases of postoperative instability or pin-site complications.

Abstract

Aims The Mako system is one of the most widely used systems for robotic arm-assisted total hip arthroplasty (THA). Two workflows for femoral preparation exists for this system – the enhanced and the express workflow. Methods We performed a retrospective cohort study comparing the accuracy of each workflow in restoring the patient’s combined offset and correcting the hip length discrepancy. The Mako derived values were compared against measured values and assessed with Bland-Altman plots and Theil’s median slope. Secondary outcomes of interest included comparison of the measured combined offset and hip length discrepancy, surgical time, incidence of postoperative instability/dislocation, as well as pin-site related complications between the two groups. Results A total of 81 patients were identified from our database: 61 in the enhanced group and 20 in the express workflow group. Bland-Altman plots demonstrated agreement between the measurements for both hip length discrepancy and combined offset. There was no significant difference in the measurement of hip length discrepancy or combined offset difference between the two groups; however, the magnitude of the latter was better in the enhanced (median 1.50 mm) compared with the express workflow group (median 3.13 mm). There was no significant difference in measured combined offset (p = 0.254), hip length discrepancy (p = 0.425), or surgical time (p = 0.548). Lastly, there were no patients with postoperative instability/dislocation nor pin-site related complications in either group. Conclusion Both techniques provide excellent outcomes with minimal risk of complications when performing a Mako robotic arm-assisted THA. Cite this article: Bone Jt Open 2026;7(2):185–194.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Cheok et al. (2026) studied this question.

synapsesocial.com/papers/698c1c53267fb587c655eb7fhttps://doi.org/10.1302/2633-1462.72.bjo-2025-0288.r1
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Accuracy of femoral component anteversion in robotic total hip arthroplasty2024 · 18 citations
  2. 2The impact of functional combined anteversion on hip range of motion: a new optimal zone to reduce risk of impingement in total hip arthroplasty2021 · 41 citations
  3. 3Projections of hip arthroplasty in OECD countries up to 20502018 · 209 citations
  4. 4Reconstruction of the hip. A mathematical approach to determine optimum geometric relationships.1979 · 357 citations
  5. 5Eine neue Methode zur r�ntgenologischen Beurteilung eines H�ftgelenkes ? Der H�ftwert1972 · 140 citations