June 1, 2018Circulation Cardiovascular Imaging396 citationsOpen Access

Diagnostic Accuracy of a Machine-Learning Approach to Coronary Computed Tomographic Angiography–Based Fractional Flow Reserve

View Full Paper
ACAdriaan CoenenYKYoung‐Hak KimMKMariusz Kruk

Key Points

  • The research aims to evaluate the diagnostic performance of a machine learning-based approach to CT-FFR compared to conventional techniques for diagnosing coronary artery disease.
  • Conducted across 5 centers in Europe, Asia, and the USA, involving 351 patients and 525 vessels with invasive FFR for comparison.
  • Calculated diagnostic performance of ML-based and CFD-based CT-FFR against invasive FFR as reference.
  • Assessed correlation and accuracy improvement using an area under curve analysis.
  • ML-based CT-FFR and CFD-based CT-FFR both had an area under curve of 0.84, significantly outperforming visual CTA with an AUC of 0.69 (P<0.0001).
  • Per-vessel diagnostic accuracy improved from 58% by CTA to 78% by ML-based CT-FFR.
  • Per-patient accuracy rose from 71% by CTA to 85% with ML-based CT-FFR, reclassifying 73% of previously false-positive results.

Abstract

Background: Coronary computed tomographic angiography (CTA) is a reliable modality to detect coronary artery disease. However, CTA generally overestimates stenosis severity compared with invasive angiography, and angiographic stenosis does not necessarily imply hemodynamic relevance when fractional flow reserve (FFR) is used as reference. CTA-based FFR (CT-FFR), using computational fluid dynamics (CFD), improves the correlation with invasive FFR results but is computationally demanding. More recently, a new machine-learning (ML) CT-FFR algorithm has been developed based on a deep learning model, which can be performed on a regular workstation. In this large multicenter cohort, the diagnostic performance ML-based CT-FFR was compared with CTA and CFD-based CT-FFR for detection of functionally obstructive coronary artery disease. Methods and Results: At 5 centers in Europe, Asia, and the United States, 351 patients, including 525 vessels with invasive FFR comparison, were included. ML-based and CFD-based CT-FFR were performed on the CTA data, and diagnostic performance was evaluated using invasive FFR as reference. Correlation between ML-based and CFD-based CT-FFR was excellent ( R =0.997). ML-based (area under curve, 0.84) and CFD-based CT-FFR (0.84) outperformed visual CTA (0.69; P <0.0001). On a per-vessel basis, diagnostic accuracy improved from 58% (95% confidence interval, 54%–63%) by CTA to 78% (75%–82%) by ML-based CT-FFR. The per-patient accuracy improved from 71% (66%–76%) by CTA to 85% (81%–89%) by adding ML-based CT-FFR as 62 of 85 (73%) false-positive CTA results could be correctly reclassified by adding ML-based CT-FFR. Conclusions: On-site CT-FFR based on ML improves the performance of CTA by correctly reclassifying hemodynamically nonsignificant stenosis and performs equally well as CFD-based CT-FFR.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Coenen et al. (2018) studied this question.

synapsesocial.com/papers/698cd14d0f5172d9365dbe92https://doi.org/10.1161/circimaging.117.007217
Ask AI
Helpful
Bookmark
Share
View Full Paper