Developing an Automatic Preparation System for Dual Clumped Isotope Measurements in Carbonates

View Full Paper
YGYangrui GuoYGYuhe GuoFTFei Tang

Key Points

  • The goal is to develop an automatic preparation system for efficient carbonate dual clumped isotope analysis.
  • Developed a fully automatic preparation system integrated with an isotope ratio mass spectrometer.
  • Utilized a Programmable Logic Controller for acid digestion and purification.
  • Tested the system with multiple carbonate reference materials.
  • Achieved high data repeatability with Δ 47 ranging from 0.003‰ to 0.023‰.
  • Maintained stability over several months with consistent normalised values.
  • Demonstrated effectiveness for carbonate-based normalisation in inter-laboratory comparisons.

Abstract

Current challenges in carbonate dual clumped isotope ( Δ 47 – Δ 48 ) analyses include the high cost of commercial systems and the low throughput associated with manual preparation across different laboratories. To address these limitations, we developed a fully automatic preparation system for carbonate samples, coupled to a ThermoFisher Scientific 253 Plus isotope ratio mass spectrometer. The system integrates a programmable logic controller with units for acid digestion, cryogenic purification, and temperature control, enabling continuous operation with minimal supervision. The flexible and low‐cost design achieved through open‐source hardware allows customisation for various research needs and can potentially be adapted for other isotope systems. The system was tested using a series of carbonate reference materials, including in‐house barium carbonates (BSD‐1, BSD‐2), ETH‐1/2/3/4, C1, C2, GU1, and ISTB‐1. The results demonstrate high data repeatability (expressed as 1 standard deviation, 1 s ), with Δ 47 ranging from 0.003‰ to 0.023‰ and Δ 48 from 0.029‰ to 0.078‰. The system's stability was validated over several months, showing no significant drift in normalised values between sessions. We also compared the Δ 47 and Δ 48 values with previously published results, suggesting the potential of carbonate‐based normalisation for achieving consistent Δ 47 – Δ 48 data across different laboratories. Future efforts should include a comprehensive inter‐laboratory comparison using equilibrated gases as primary anchors alongside a wider range of carbonate reference materials to rigorously establish measurement reproducibility across laboratories.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Guo et al. (2026) studied this question.

synapsesocial.com/papers/69926503eb1f82dc367a0e33https://doi.org/10.1111/ggr.70041
Ask AI
Helpful
Bookmark
Share
View Full Paper