Continuous coagulation monitoring in human blood samples via magnetic particle spectroscopy

MBMaria-Josephina BuhnéUniversity Hospital Schleswig-HolsteinJAJustin AckersMAMandy AhlborgFraunhofer Research Institution for Individualized and Cell-Based Medical Engineering

Key Points

  • To assess the feasibility of using magnetic particle spectroscopy (MPS) for continuous blood coagulation monitoring.
  • Blood samples mixed with Resotran were analyzed in a custom MPS system for 45 minutes.
  • The fifth harmonic signal amplitude was measured under different conditions, including anticoagulant presence and mechanical stress.
  • Temperature was monitored to exclude thermal effects during measurements.
  • All blood samples with Resotran showed a signal decrease over time.
  • Samples with anticoagulants exhibited minimal or no signal decrease compared to control samples.
  • Mechanical stress resulted in signal decay across all samples, linking signal changes to coagulation.

Abstract

Introduction: Magnetic Particle Imaging (MPI) is a radiation-free imaging modality based on the nonlinear magnetic response of iron oxide nanoparticles, providing high sensitivity and real-time, quantitative, background-free imaging. With the clinical approval of Resotran as an MPI-suitable tracer and the development of first human-scale scanners, clinical applications are within reach. Magnetic Particle Spectroscopy (MPS), the non-imaging counterpart of MPI, enables sensitive analytics by exploiting the signal response of magnetic nanoparticles. In this pilot study, we prove the potential of MPS to continuously monitor blood coagulation in real time. Methods: Blood samples from five volunteers were mixed with the commercial magnetic resonance imaging contrast agent Resotran. The dynamics of the particle signal were assessed in a custom-built MPS-system for a duration of 45 minutes under various conditions, including the presence of anticoagulants (EDTA, Heparin, Citrate) and mechanical stress. The signal amplitude of the fifth harmonic of the MPS was analyzed. To exclude potential thermal effects, the temperature inside the MPS was monitored by using a fiber optic thermometer during the measurements. Results: All Resotran-containing blood samples showed a signal decrease over time. Samples with anticoagulants exhibited no relevant signal decrease (EDTA, Citrate) or a smaller decrease (Heparin) compared to samples without anticoagulants. Additionally, mechanical stress induced a signal decay in all samples, further indicating the link between the observed MPS signal decay and blood coagulation. Conclusion: This study shows that continuous monitoring of human blood coagulation via MPS is feasible, making bedside coagulation monitoring in clinical settings a concrete perspective.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Buhné et al. (2026) studied this question.

synapsesocial.com/papers/69aa70c8531e4c4a9ff5aea8https://doi.org/10.15480/882.16803