Transcutaneous electrical nerve stimulation (TENS) is constrained by high skin impedance and unstable electrode contact. This study proposes a novel composite electrode system comprising a polyvinyl alcohol/silver (PVA/Ag) microneedle array and a highly conductive polyacrylamide/lithium chloride (PAAm/LiCl) hydrogel. The PVA/Ag microneedles (~365 µm in height, ~48 µm tip diameter) possess sufficient mechanical strength to penetrate the stratum corneum, establishing a low-resistance pathway. The complementary PAAm/LiCl hydrogel exhibits high conductivity (10.28 S/m) and mechanical flexibility, further optimizing the interface contact. The experimental results demonstrate that this composite system achieves low electrochemical impedance and induces stable, clear electromyographic responses in vivo. It effectively addresses the common issues of electrode detachment and signal attenuation associated with conventional electrodes, offering a promising hardware solution for efficient and comfortable wearable rehabilitation devices.