This report demonstrates a technique for recording monocular VEPs under binocular conditions, indicating clinical utility in VEP testing.
This report describes a technique for recording the visual evoked potential (VEP) arising from 1 eye under conditions of binocular pattern stimulation. The subject views a sinusoidally modulated vectographic checkerboard with orthogonally oriented polarizing filters before the eyes. This technique results in waveforms that contain monocular VEPs distributed in time. The technique is sensitive to interocular suppression and controls for the influence of various extraneous sources of variability. It is useful in clinical VEP testing.
No takes yet. Share an insight, caveat, or question.
May et al. (1978) studied this question.