March 21, 20260 citationsOpen Access

Normalized Quantitative Assessment of Receive-Coil Sensitivity using MPI Transfer Function Measurements

View Full Paper
FTFlorian ThiebenUniversität HamburgEKEgor KretovFraunhofer Research Institution for Individualized and Cell-Based Medical EngineeringOTOnurcan Topcu

Key Points

  • The aim is to develop an efficient method to measure receive-coil sensitivity for MPI applications.
  • Utilized a calibration coil for measurements
  • Conducted a small number of targeted assessments
  • Analyzed field inhomogeneities during imaging
  • Demonstrated a new approach for measuring normalized receive-field profiles
  • Achieved improved accuracy in system matrix generation
  • Highlighted the efficiency of fewer measurements for reliable data

Abstract

In human-scale magnetic particle imaging (MPI), field inhomogeneities arise because a large fraction of the bore volume is utilized for imaging. Accurate knowledge of the receive-coil sensitivity is essential for hybrid system-matrix generation and model-based system matrix reconstruction. In this work, we present an efficient approach for measuring normalized receive-field profiles using a calibration coil and a small number of measurements.