Key result
Jugular access for leadless pacemakers matches femoral electrical performance while reducing fluoroscopy and access complications.
Why the study?
Comparative evidence between transjugular and transfemoral venous access for leadless pacemaker implantation remains limited.
Does the transjugular approach reduce complications or improve procedural efficiency compared to transfemoral access in patients undergoing leadless pacemaker implantation?
Meta-Analysis
Does the transjugular approach reduce complications or improve procedural efficiency compared to transfemoral access in patients undergoing leadless pacemaker implantation?
The transjugular approach for leadless pacemaker implantation offers comparable electrical performance to the standard transfemoral approach while significantly reducing fluoroscopy exposure and access-related complications.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Validates jugular access as a safer alternative for leadless pacemaker implantation; extends.
BACKGROUND: Leadless pacemakers (LPMs) have been shown to reduce complications compared with conventional transvenous pacing. Transfemoral venous access (FA) is the standard implantation approach; however, the transjugular approach (JA) has recently emerged as an alternative strategy, but comparative evidence remains limited. Therefore, we perform a systematic review and meta-analysis comparing JA and FA for LPMs implantation in terms of electrical performance, procedural efficiency, and complications. METHODS: All studies comparing JA and FA for LPMs were systematically identified. Pooled analyses were performed for pacing threshold, sensed R-wave amplitude, impedance, total procedural time, fluoroscopy duration, and procedural complications. RESULTS: Three observational studies were included. No significant differences were observed between FA and JA in mean pacing threshold (MD 0.05 ms; 95% CI -0.03 to 0.12), mean sensed R-wave amplitude (MD 0.31 mV; 95% CI -0.51 to 1.14), or impedance (MD 45.79 Ω; 95% CI -27.03 to 118.61). Total procedural time was comparable between approaches (MD -0.44 min; 95% CI -8.47 to 7.60). JA was associated with a significant reduction in fluoroscopy time compared with FA (MD 2.73 min; 95% CI 0.66 to 4.81). JA also significantly reduced overall procedural complications (1.8% vs. 15%; OR 0.12, 95% CI 0.04 to 0.39) and wound care-related adverse events (0% vs. 14%; OR 0.07, 95% CI 0.01 to 0.36). No differences were observed in device dislodgement, pericardial effusion, or emergent device removal. CONCLUSIONS: JA provides comparable electrical performance to FA while reducing fluoroscopy exposure and access-related complications, supporting its use as a safe alternative for LPMs implantation.
Giannino et al. (2026) conducted a meta-analysis in Leadless pacemaker implantation. Transjugular venous access vs. Transfemoral venous access was evaluated on Overall procedural complications (OR 0.12, 95% CI 0.04 to 0.39). Transjugular access for leadless pacemaker implantation significantly reduced overall procedural complications compared with transfemoral access (1.8% vs. 15%; OR 0.12, 95% CI 0.04-0.39).