Advancing the Reproducibility and Repeatability of Capillary Zone Electrophoresis-Mass Spectrometry-Based Top-Down Proteomics by an Improved Capillary Coating Procedure

View Full Paper
YYYifan YueFFFei FangGGGuangyao Gao

Key Points

  • To enhance the reproducibility and repeatability of capillary zone electrophoresis-mass spectrometry (CZE-MS) in top-down proteomics through an improved capillary coating procedure.
  • Developed an improved linear polyacrylamide (LPA) coating procedure for CZE-MS.
  • Focused on optimizing the degassing step of the polymerization solution.
  • Conducted 210 CZE-MS runs on three protein samples across 200 hours using two mass spectrometers.
  • Evaluated the variability of migration times and proteoform intensities.
  • Achieved relative standard deviations (RSDs) of less than 2.6% for migration times across numerous runs.
  • Identified 274 ± 20 proteoforms and 146 ± 10 proteins in HeLa cell lysates with consistent intensities over multiple runs.
  • Demonstrated the feasibility of the coating procedure for use by other research groups.

Abstract

Capillary zone electrophoresis (CZE)-mass spectrometry (MS) has attracted tremendous attention in top-down proteomics (TDP). However, its reproducibility and long-term repeatability for TDP remain concerns, most likely due to capillary coating. Here, we present an improved procedure for making linear polyacrylamide (LPA) coating, the most widely used coating in CE-MS-based proteomics, to boost the reproducibility and long-term repeatability of CZE-MS-based TDP. We focused on the step of degassing the polymerization solution, a critical step for achieving consistent LPA coating quality. The CZE-MS system using LPA-coated capillaries prepared with the optimal degassing procedure produced excellent reproducibility and repeatability for proteoform analysis. The 210 CZE-MS runs of three protein samples (a standard protein mixture, an E. coli cell lysate, and a HeLa cell lysate) across 200 h of instrument time with two MS instruments demonstrated the reliability of our conclusion. The optimal condition produced relative standard deviations (RSDs) of less than 2.6% in the migration time of proteoforms across dozens of CZE-MS runs without migration time alignment. CZE-MS/MS analysis of a HeLa whole cell lysate identified 274 ± 20 proteoforms (RSD ∼ 7%) and 146 ± 10 proteins (RSD ∼ 7%) across 10 successive runs and yielded consistent proteoform intensities. The coating procedure can be easily adopted by other research groups, as evidenced by our 2025 CE-MS summer school data. All results demonstrate that CZE-MS using the optimal LPA-coating procedure is ready for broad adoption to enable reproducible and repeatable measurements of proteoforms in complex samples.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Yue et al. (2026) studied this question.

synapsesocial.com/papers/69ccb7b016edfba7beb89bc4https://doi.org/10.1021/acs.jproteome.5c01194
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Pilot Evaluation of the Long-Term Reproducibility of Capillary Zone Electrophoresis–Tandem Mass Spectrometry for Top-Down Proteomics of a Complex Proteome Sample2024 · 13 citations
  2. 2Selective Comprehensive Online NanoLCxCZE-MS Platform for Top-Down Proteoform Analysis.2025
  3. 3Recent Advances (2023–2025) of Capillary Electrophoresis‐Mass Spectrometry (CE‐MS) for Top‐Down Proteomics2026 · 2 citations
  4. 4Solid-Phase Microextraction-Aided Capillary Zone Electrophoresis-Mass Spectrometry: Toward Bottom-Up Proteomics of Single Human Cells2024 · 16 citations
  5. 5Top-Down Proteomics Analysis of Picogram-Level Complex Samples Using Spray-Capillary-Based Capillary Electrophoresis–Mass Spectrometry2024 · 9 citations