Zero-fluoroscopy pulsed field ablation proves feasible for paroxysmal AF in small case series.
Why the study?
Zero-fluoroscopic pulsed field ablation (PFA) represents an advancement in pulmonary vein isolation for atrial fibrillation, but its workflow feasibility was not established.
Is a zero-fluoroscopic pulsed field ablation workflow feasible in patients with paroxysmal atrial fibrillation?
Observational (n=15)
Is a zero-fluoroscopic pulsed field ablation workflow feasible in patients with paroxysmal atrial fibrillation?
Answers are grounded in this paper and cite their sources.
A zero-fluoroscopic workflow for pulsed field ablation using intracardiac echocardiography and a visualized sheath is feasible for pulmonary vein isolation in paroxysmal atrial fibrillation.
Background: Zero-fluoroscopic pulsed field ablation (PFA) is the next step in advancing pulmonary vein isolation (PVI) for atrial fibrillation (AF). Methods: A workflow incorporating a variable-loop circular PFA catheter, intracardiac echocardiography (ICE), and a visualized sheath was applied to 15 paroxysmal AF patients. Results: It consisted of three steps: (1) positioning a non-magnetic mapping catheter in the coronary sinus after creating a right atrial matrix, (2) transseptal puncture with the sheath advancement to the left atrium under ICE guidance, and (3) PFA confirming catheter-tissue contact with tissue proximity indication. Conclusions: This zero-fluoro PFA workflow is feasible without advanced technique.
No takes yet. Share an insight, caveat, or question.
Hirata et al. (2024) conducted an observational in Paroxysmal atrial fibrillation (n=15). Zero-fluoroscopic pulsed field ablation (PFA) workflow was evaluated on Feasibility. A zero-fluoroscopic pulsed field ablation workflow was feasible in 15 patients with paroxysmal atrial fibrillation.