A comparison of the dose data acquired in this study reveals that the MC algorithm calculations are closer to the 60 Gy prescribed dose for PTV, while the difference between the PB and MC algorithms was found to be non-significant. Because of the major difference arising from the dose calculation techniques by TPS that was observed in the MLD with significant medium heterogeneity, we recommend the use of the MC algorithm in such heterogeneous sites.