January 9, 2020International Journal of Radiation Oncology*Biology*Physics479 citationsOpen Access

Design, Implementation, and in Vivo Validation of a Novel Proton FLASH Radiation Therapy System

View Full Paper
EDEric S. DiffenderferUniversity of PennsylvaniaIVIoannis I. VerginadisUniversity of PennsylvaniaMKMichele M. KimUniversity of Pennsylvania

Key Points

Key points are not available for this paper at this time.

Abstract

We have designed and dosimetrically validated a FLASH-PRT system with accurate control of beam flux on a millisecond time scale and online monitoring of the integral and dose delivery time structure. Using this system, we found that FLASH-PRT decreases acute cell loss and late fibrosis after whole-abdomen and focal intestinal RT, whereas tumor growth inhibition is preserved between the 2 modalities.