Answers are grounded in this paper and cite their sources.
To determine the effect of peripherally fenestrated contact lenses on central corneal edema, 6 subjects (12 eyes) wore steeply fitted lenses of identical dimensions but with various numbers of 0.1-mm diameter peripheral fenestrations for 3-hr periods. At the end of the wearing periods, thickening, steepening, and central clouding of the cornea were generally reduced as the number of fenestrations was increased. However, 20 fenestrations were required before the edema was reduced to a clinically acceptable level.
No takes yet. Share an insight, caveat, or question.
Harris et al. (1977) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: