A fast and simple technique composed of dispersive liquid-liquid microextraction (DLLME) and online preconcentration MEKC with diode array detection was developed for the determination of four phenoxyacetic acids, 2,4,5-trichlorophenoxyacetic acid, 2,4-dichlorophenoxyacetic acid, 2,6-dichlorophenoxyacetic acid, and 4-chlorophenoxyacetic acid, in drinking water. The four phenoxyacetic acids were separated in reversed-migration MEKC to the baseline. About 145-fold increases in detection sensitivity were observed with online concentration strategy, compared with standard hydrodynamic injection (5 s at 25 mbar pressure). LODs ranged from 0.002 to 0.005 mg/L using only the online preconcentration procedures without any offline concentration of the extract. A DLLME procedure was used in combination with the proposed online preconcentration strategies, which achieved the determination of analytes at limits of quantification ranging from 0.2 to 0.5 μg/kg, which is far lower than the maximum residue limits established by China. The satisfactory recoveries obtained by DLMME spiked at two levels ranged from 67.2 to 99.4% with RSD <15%, making this proposed method suitable for the determination of phenoxyacetic acids in water samples.