Hydrogen-deuterium exchange mass spectrometry (HX-MS) has made important contributions to the study of protein structure and function. Unfortunately, it is not known for low limits of detection, when compared with other forms of peptide-based or bottom-up protein MS methods. Systems perform poorly on sub-pmol quantities of protein states with greater than 300 kDa of unique sequences. The HX-MS analysis of complex protein states would be possible if proteomics-grade configurations could be used reliably, but temperature and temporal constraints have proven to be significant design challenges. Here, we describe an integrated HX-MS ion source operating on a vented-column geometry, which brings regulated column cooling right to the spray tip. The design offers chromatographic peak widths of 2-6 s (FWHM). It provides stable operation at 500 nL min⁻¹, while retaining deuteration levels comparable to conventional geometries. We demonstrate at least a 50-fold improvement in protein consumption levels, and illustrate robustness by measuring peptide-averaged protection factors for 90% of DNA-PKcs, a 469 kDa protein, from 0.5 pmol injections.
No takes yet. Share an insight, caveat, or question.
Sheff et al. (2017) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: