A sensitive, accurate and highly stereoselective assay for the simultaneous determination of venlafaxine (VEN) and its equipotent metabolite, O-desmethyl venlafaxine (ODV), in human plasma was developed and validated. Analytes were simultaneously extracted from plasma using solid-phase extraction and detected by tandem mass spectrometry in positive ion mode with a turbo ion spray interface. Deuterium-labeled VEN and ODV were used as internal standards. Chromatographic separation was performed on a Chiral AGP column, using a time programmed gradient flow with a total run time of 16 min. The method has a lower limit of quantitation of 0.60 ng/mL. The assay was linear over a range 0.60-300.00 ng/mL for both the enantiomers of VEN and ODV, respectively, with coefficient of correlation > 0.99. The extraction recoveries were >77.0% on an average for all the four analytes. The analytes were found stable in plasma through three freeze (-15 °C) and thaw cycles and under storage at room temperature for 8 h, and also in mobile phase at 10 °C for 54 h. The method has shown good reproducibility, with intra- and inter-day variation coefficients < 9%, for all the analytes, and has proved to be very reliable for analysis of VEN and its metabolite in clinical study samples.
No takes yet. Share an insight, caveat, or question.
Dutta et al. (2012) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: