PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
July 1, 1988Anesthesia & Analgesia174 citations

The Effect of Nail Polish on Pulse Oximetry

View Full Paper
CCCharles J. CotEGEsther GoldsteinWFWilliam H. Fuchsman

Key Points

  • To evaluate how various nail polish colors affect the accuracy of oxygen saturation measurements obtained via pulse oximetry.
  • Randomized, blind study in 14 adult volunteers who had blue, green, purple, black, and red nail polish applied to their fingernails under room air conditions.
  • Oxygen saturation was recorded with a Nellcor N100 pulse oximeter and interpreted in a blinded fashion.
  • Spectrophotometry was performed to determine the absorption spectra of the applied polishes, nine additional polishes, and three intravenous dyes at 660 nm and 940 nm.
  • Black, blue, and green nail polish significantly lowered pulse oximetry readings, with blue and green producing greater decreases than purple and red.
  • Artifactual desaturation correlated strongly with the difference between light absorbance at 660 nm and 940 nm (r = 0.95).
  • Spectrophotometric profiles accurately predicted that brown-red polish causes significant optical interference during pulse oximetry monitoring.

Abstract

A randomized, blind study examined the effect of nail polish color on measurement of oxygen saturation by pulse oximetry. Fourteen adult volunteers had blue, green, purple, black, and red nail polish applied to their finger nails. A strip-chart recording of oxygen saturation (Nellcor N100) was made in room air and later interpreted in a blinded fashion. The absorption spectra of the five polishes were determined by spectrophotometry. The spectra of nine other nail polishes and three intravenous dyes also were examined. Black, blue, and green nail polish significantly lowered oximeter readings of oxygen saturation. Blue and green produced greater decreases than purple and red; black produced an intermediate decrease. Some but not all nail polishes absorbed light at the wavelengths used by the pulse oximeter (660 nm and 940 nm). The degree of artifactual desaturation correlated best with the difference between absorbance at 660 nm and absorbance at 940 nm (r = 0.95). Spectrophotometric absorbance data suggest that other colors may interfere with pulse oximetry. On the basis of spectrophotometric data, brown-red nail polish was predicted to interfere with oximetry; subsequent pulse oximetry measurements confirmed the prediction. Nail polish should be removed routinely before pulse oximetry monitoring.