Variations in use of a commercially available pressure monitor exist, and errors are common. Proper use improved accuracy, but even with proper technique, 40% of the measurements were >5 mm Hg from the actual pressure. Based on our data, measurement accuracy with this device should be questioned and viewed within a range. Regular review and education of technique is strongly recommended.
No takes yet. Share an insight, caveat, or question.
Large et al. (2015) studied this question.