Answers are grounded in this paper and cite their sources.
A mechanistic explanation for previously observed safety improvements with microprocessor-controlled prosthetic knees is needed. A repeated measures design of 15 subjects with unilateral transfemoral amputation was used to assess changes between baseline use of their standard of care, mechanical pros-theses, and a C-Leg microprocessor-controlled prosthetic knee. The primary outcome measures were sensory dependency scores for somatosensory, visual, vestibular, and visual preference, which were calculated based on a Sensory Organization Test. Falls during posturographic assessment were also recorded. Somatosensory system dependency significantly increased (p = 0.047) while using the C-Leg compared to a nonmicroprocessor prosthetic knee (NMPK). Reliance on visual with vestibular input and reliance on vestibular input alone were not significantly increased with C-Leg use (p = 0.41 and p = 0.15, respectively). When utilizing the C-Leg, there was a significant reduction in the average number of falls (p = 0.03). Hence, increased reliance on somatosensory input is a possible explanation for improved balance with use of a microprocessor prosthetic knee (MPK).
No takes yet. Share an insight, caveat, or question.
Highsmith et al. (2014) studied this question.